Low Differential Pressure Transmitter Dwyer Series 610
Low Differential Pressure Transmitter Dwyer Series 610
-
The Series 610 Low Differential Pressure Transmitters are capable of measuring the pressures and flow of air or non-conducting gases at high resolutions.
-
Designed specifically for clean rooms, isolation rooms, and other critical environments, the Series 610 is ideal for situations when accurate and reliable pressure monitoring is essential.
-
Easy access pressure ports and electrical connections, removable process heads, and detachable terminal blocks make for fast and easy installation.
-
The Series 610 transmitters are available for air pressure ranges as low as 0.1 in w.c. full-scale.
-
The Series 610 transmitters can be ordered as either ±0.25% or ±0.5% FS accuracy, base mount or din rail mount and the option of a digital read out display.
-
In addition, a calibration key can be ordered that allows the user to set zero and span. One key will work on multiple transmitters.
Applications:
-
Lab Fume Hood Control
-
Clean Room Applications
-
Flow Measurements & Controls
-
Filter Monitoring
-
Furnace Draft Measurement
-
Process Control
Specifications:
-
Service:Air or similar non-conducting gases.
-
Accuracy:±0.25% or ±0.5% F.S.
-
Temperature Limits:-20 to 160°F (-29 to 71°C).
-
Pressure Limits:2 psi (0.137 bar).
-
Power Requirements:13.5 to 30 VDC.
-
Loop Resistance:800 ohms max.
-
Current Consumption:25 mA.
-
Zero and Span Adjustments:External security key pendant.
-
Response Time:0.02 to 0.04 seconds.
-
Electrical Connections:Detachable screw terminal connector.
-
Process Connections:3/16" O.D. barbed brass fittings on removable process head.
-
Enclosure Rating:Fire retardant ABS.
-
Mounting Orientation:Vertical.