A  pressure sensor process shall consider the following:
1 , to prevent contact with the transmitter and the corrosive medium or hot;
2,  to prevent garbage deposited in the conduit;
3,  when measuring the liquid pressure, the pressure taps should be opened in the process side of the line, in order to avoid precipitate product bagasse;
4 , the measurement of gas pressure, the pressure taps should be opened at the top of the flow line, and the transmitter should be installed in the upper flow of the pipeline, so that the accumulation of liquid flow easily injected the pipeline;
5 , the guide pressure pipe should be installed in a small temperature fluctuations place;
6, steam or other high temperature measuring media, must be connected to add buffer tube (coil), and other condensers, should not exceed the limit temperature transmitter;
7, when the winter frost occurs, the transmitter is installed outdoors freezing measures must be taken to prevent the liquid inside the mouth due to the impulse ice volume expansion, leading to the sensor is damaged;
8, when measuring the liquid pressure, the position of the transmitter should be installed to avoid the impact of the liquid (water hammer phenomenon), so as not to damage the sensor overvoltage;

9, the wiring, the cables through the waterproof connector (accessory) or about the nature pipe and tighten the nut tightly sealed to prevent rainwater seepage into the inner transmitter housing through cable.


Specification:

 

Name

WP1151 Capacitive Smart Differential Pressure Transmitter

Measurement range

0~0.1kPa to 0~40MPa

Measurement type

GP, DR, DP, HP, AP, DP/GP

Power supply

12~45V DC, normally 24V DC

Output signal

4-20mA DC/ HART protocol

Span and zero point

Adjustable

Accuracy

0.25%FS

Humidity

Relative humidity 5~95%

Diaphragm

Stainless steel 316 / Monel/  Hastealooy C / Tantalum

Media

liquid, gas, or steam flow

Explosion proof

dIIBT4, iaIICT6

Overload

1.15 times

Filled liquid

Silicone oil

Pressure Range’s Ordering codes

1E

0~25-157Pa

6E

0~117-690 KPa

2E

0~0.125-1.5KPa

7E

0~345-2068 KPa

3E

0~1.3-7.5KPa

8E

0~1170-6890 KPa

4E

0~6.2-37.4 KPa

9E

0~3450-20680 KPa

5E

0~31.1-186.8 KPa

0E

0~6890-41370 KPa

More technical parameters, please feel free to contact with us.


About Us

About Us

Xi`an Ao Xin Automation Instrument Co., Ltd., is located in Xi`an city which is the authoritive centre of instrument and gauge .It is cooperated with many related domestic cooperation to research ,sale and service and also obtained the cert of ISO9001-2008. overall strength of the industry 's leading domestic well-known enterprises .Over the years, the company under the "integrity, innovation , development and harmony " principle of service .

Contact Us

Website:http://www.aosoonyb.com
TEL:+86-029-86691808
                   +86-029-84381810
    Email:xaaxyb@126.com          Email:wsb2828@126.com

Top
点击这里给我发消息 xaqksm wang01-Send Email  Amber
 E-mail:wsb2828@126.com
My status
My status